Test Items | 1. ABO Forward Blood typing test; 2. ABO Reverse blood typing test; 3. Rh (D) blood typing test 4. Anti-human globulin assay |
Test Carrier | Column Agglutination Cards and Gel Cards |
Processing Rate | ABO Forward/Reverse Blood Grouping Card process 96 cards per hour |
Specimen Capacity | 144 profiles |
Reagent Capacity | 14 Reagent Positions, 12 with Automatic Mixing Function |
Volume | 96 column agglutination cards |
Barcode Scanning | Sample Scan and Test Card Scan |
Loading Arm | 1 sample loading arm, 2 sample loading channels, using disposable tip, with liquid level detection function, missing needle detection (detecting whether the tip is loaded), clot detection |
Robotic Arm | 1 robotic arm for transferring reagent cards |
Sampling repeatability and accuracy | Sampling volume (ul) Repeatability CV (%) Accuracy (%) |
10 ≤3% ≤±5% | |
50 ≤2% ≤±1% | |
100 ≤1% ≤±1% | |
Liquid Adding Capacity | 5ul~800ul |
incubator | 18Cards |
Incubation Temperature | Room Temperature~60℃ |
Centrifuge | 2*12 Cards, Maximum Speed 3000r/min |
Interpretation Module | Composed of CMOS and backlight source, the experimental results are automatically interpreted, and the report is issued |
Operation Module | Support continuous loading of specimens, reagents, blood cards, and cyclic sample loading |
Connector | Power cord and USB |
Operating system and software | Chinese WindowsXP/7/8 Operating System, Access Database, Ability to interface with the hospital's LIS System |
Operating environment | Temperature: 15℃-30℃ Humidity: 30%-85% |
Power Source | 220±10% V,50/60Hz |
Size | 1320(L)*830(W)*1710(H) |
Applicable to medical units with monthly testing volume ≥ 2000 reagent cards |
Copyright © 2011 - 2017 Dexiang Biotech Co.,LTD
备案号:津ICP备14003466号-1 互联网药品信息服务备案号:2024010012 ICP Number 12019002000261